Face Mask Particulate Filtration Efficiency Tester

如果您对该产品感兴趣的话,可以
产品名称: Face Mask Particulate Filtration Efficiency Tester
产品型号: AG-2663B
产品展商: SHENZHEN ANGUI OF CHINA
产品文档: 无相关文档

简单介绍

(Configure photometer imported from the United States, with loading test) (Test for salinity or oily aerosols)


Face Mask Particulate Filtration Efficiency Tester  的详细介绍

A. Principle:

Scattering photometer, analyze and detect the concentration ratio of the oily aerosol before and after the filter material, calculate the penetration rate and filtration efficiency. Using the scattering photometer to measure the concentration of paraffin oil aerosol according to the latest international standard EN 149.


B. Scope of Application

It is used to test the particulates filtration efficiency of daily face masks and filter materials.


C. Applicable Standards:

GB 2626-2019: Respiratory protection—Non-powered air-purifying particle respirator

GB/T 32610-2016: Technical specification of daily protective mask

GB/T19083-2010: Technical requirements for protective face mask for medical use

YY0469-2011: Surgical mask

EN 149-2019

BSEN13274-7:2019

D. Technical Parameters

1.     Particle Diameter: mass concentration median diameter of 0.6 microns

2.    Flow Range: (20100) L/min, accuracy 2%

3.   Filtration Efficiency: (0.00199.999)% 

4.    Resistance: 01000Pa, accuracy 2%

5.     Test Area: 100cm² or customized fixture

6.     Aerosol: oil (paraffin oil)

7.    Aerosol Concentration: (0.1200) mg/m³ (oily)

8.     Test Mode: filtration efficiency based on the different load of aerosol

9.     Test Temperature: (25±5)℃

10.  Relative Humidity: (30±10)%

11.  Dimensions: 1600*600*2300 (mm)

12.  Weight: 90kg

13.  Power Supply: AC220V 50Hz

14.  Compressed Air: 550kPa, 198 std. L/min 

E System Composition

1、Aerosol

  Equipped with a set of the independent aerosol generator; oily particulate generator


2、Air-Flow

Consists of a vacuum pump, air-flow controller, mass flow transmitter, differential pressure transmitter, and air pipe.


3、Photometer

Using the 90-degree right angle light scattering principle, the built-in air pump is used to draw aerosol particles into the optical chamber, and the mass concentration of the particles is measured by the amount of light scattered. Dynamic measurement range: can measure the mass density of ions below 1.0 microns, below 2.5 microns, below 4.0 microns, and below 10.0 microns. Detection range: 0.001~400mg/m³.

The photometer at the front of the filter material sample measures the mass concentration at the front; the photometer at the back of the filter material sample measures the mass concentration of the particles at the back. The ratio of the mass concentration at the front and back is used to calculate the filtration efficiency of the filter material sample.


4、Control System

PLC control measures airflow and pressure difference before and after filtration.

Calculate the particle concentration filtration efficiency and mass concentration filtration efficiency of different particle sizes when the loading capacity is in the range of 30mg-400mg.


Print out measurement output: ( loading capacity can be set as: 1mg----30mg----400mg)


-------------------------------------------------------------------------

Rate

Differential Pressure

Aerosol Particles C

Loading Capacity (mg)            Filtration Efficiency %

1.0                                                                  97

5.0                                                                 96

10.                                                               96

30                                                               92

100                                                              71

200                                                              62

400                                                             35

------------------------------------------------------------------------------